Seiko TG/DTA 6200

Seiko Instruments TG/DTA 6200 User Information

Temperature Range Ambient – 1500°C
Automatic Cooling Unit Forced Air Cooling
Cooling Time 1000°C – 50°C within 15 min
Balance Type Horizontal Differential Type
TG measurement Range 200 mg
TG Sensitivity 0.2 µg
DTA Measurement Range =/- 1000 µV
DTA Sensitivity 0.06 µV
Programmable Rate 0.01 – 20 °C/min
Sample Pan Material Platinum, Aluminum
Sample Pan Volume 45 µl
Analysis Atmosphere Air, Nitrogen (Helium upon request)
Gas Flow Rate 0 – 1000 ml/min